Explore how X-ray diffraction and thermal analysis support the characterization of metal-organic frameworks and covalent organic frameworks.
Metal-organic frameworks (MOFs) and covalent organic frameworks (COFs) are promising materials for gas storage, carbon capture, catalysis, sensing and energy storage. Understanding their structure and physicochemical behaviour is essential for evaluating their suitability for these applications.
But how can researchers determine the crystal structure of very small MOF crystals? And how can they measure gas adsorption and desorption while observing changes in the material?
Two Rigaku application notes illustrate complementary analytical approaches. The first demonstrates crystal structure determination using benchtop powder X-ray diffraction (XRD). The second investigates CO₂ and water adsorption using simultaneous thermal analysis (STA) with sample observation.
Discover how these techniques can contribute to your material characterization workflow.
Why is MOF and COF characterization important?
MOFs are porous crystalline materials composed of metal ions or clusters connected by organic linkers. COFs consist of organic building blocks connected through covalent bonds.
Both material classes attract considerable research interest because their structures and properties can be tailored for specific applications.
Characterization helps researchers investigate important parameters, including:
Crystal structure and phase identification: Understanding the structural arrangement and crystalline phases present.
Thermal stability: Determining how materials respond to temperature changes.
Gas adsorption and desorption: Evaluating mass changes associated with gas uptake and release.
Environmental response: Investigating how materials behave under different temperatures, gas atmospheres and humidity conditions.
No single analytical technique provides all this information. Combining complementary methods allows researchers to build a more comprehensive understanding of their materials.
Crystal structure determination of MOFs using benchtop XRD
Overcoming the challenges of small MOF crystals
Determining the crystal structure of newly synthesized MOFs can be challenging. Many MOFs crystallize as particles smaller than 1 µm, making conventional single-crystal X-ray diffraction difficult.
Powder X-ray diffraction offers an alternative for investigating crystalline materials when suitable single crystals are unavailable.
In Rigaku application note XRD1014, researchers used the Rigaku MiniFlex 600 benchtop X-ray diffractometer to determine the crystal structure of a copper-based MOF directly from powder diffraction data.
Application example: Ab-initio structure determination of Cu(ta)₂
The Cu-MOF powder was measured using the MiniFlex 600 equipped with a high-speed one-dimensional detector.
The experimental conditions included:
- Reflection geometry with a silicon zero-background sample holder.
- A diffraction range of 5–110° 2θ.
- An angular step width of 0.02°.
- A scanning speed of 5°/min.
- A total data acquisition time of just 22 minutes.
The resulting diffraction pattern was indexed using N-TREOR. This identified the tetragonal space group I4₁/amd (No. 141).
Following pattern decomposition, researchers obtained an initial structure model using the charge-flipping method. Rietveld refinement was then applied to refine structural parameters, including unit-cell dimensions and atomic coordinates.
This application demonstrates that a benchtop XRD system can provide data suitable for ab-initio crystal structure determination in a challenging MOF powder sample.
The feasibility of this approach depends on factors such as sample crystallinity, diffraction quality and structural complexity.
Instrument spotlight: Rigaku MiniFlex 600
The Rigaku MiniFlex 600 is a benchtop powder X-ray diffractometer designed for crystalline materials analysis.
Key features:
- 600 W sealed X-ray tube.
- Compact benchtop configuration.
- No external chiller required.
- SmartLab Studio II software for measurement and analysis.
- Powder diffraction applications including phase identification and structural analysis.
Evaluating CO₂ and water adsorption in MOFs using thermal analysis
Understanding gas uptake and release
Beyond crystal structure, adsorption behaviour is an important property of MOFs developed for gas capture, storage and separation.
Researchers need to understand how materials interact with different gases and how these interactions change under controlled conditions.
Thermogravimetric analysis (TGA) measures changes in sample mass. When combined with differential scanning calorimetry (DSC), simultaneous thermal analysis provides complementary thermal information.
A sample observation STA configuration adds visual monitoring, allowing researchers to correlate mass changes with changes in the sample’s appearance.
AddSPX proudly supports this mission by providing Rigaku’s advanced analytical technologies to laboratories and recycling facilities across the BeNeLux.
Discover how Rigaku XRF solutions can help your recycling process achieve new levels of precision and sustainability.
Application example: CO₂ and H₂O adsorption in HKUST-1
Rigaku application note B-TA1042 investigates the adsorption behaviour of HKUST-1, a copper-based MOF.
Researchers used the Rigaku Sample Observation STAvesta, combined with humidity control, to evaluate CO₂ and water adsorption under controlled atmospheres.
After pretreatment at 180°C to remove adsorbed substances, the sample was cooled to room temperature.
Two adsorption experiments were performed.
CO₂ adsorption
When the atmosphere changed from helium to CO₂, the sample showed an 11.7% mass increase due to CO₂ adsorption. Returning the atmosphere to helium resulted in rapid CO₂ desorption.
Water adsorption
In a separate experiment, changing from dry nitrogen to humidity-controlled nitrogen at 25°C and 80% relative humidity produced a 33.2% mass increase.
The subsequent return to dry conditions caused a more gradual mass loss through dehydration.
Combining mass measurements with visual observation
Sample observation revealed another interesting difference between the adsorption processes.
During CO₂ adsorption, HKUST-1 retained its dark blue appearance. During water adsorption, however, the sample changed to light blue.
The researchers attributed this colour change to probable hydration of the Cu²⁺ sites.
Combining these observations with gravimetric data helped distinguish the material’s responses to CO₂ and water exposure.
The study demonstrates how sample observation thermal analysis can investigate adsorption amounts, adsorption and desorption behaviour, and visible material changes within the same experiment.
Instrument spotlight: Rigaku STAvesta
The Rigaku STAvesta combines thermogravimetric analysis and differential scanning calorimetry for advanced thermal characterization.
Key features:
- Simultaneous TG-DSC measurements.
- Temperature range from room temperature to 1100°C, with extended options.
- Optional 52-position autosampler.
- Dynamic DSC capabilities.
- Sample observation and humidity-control configurations for specialized applications.
Why combine XRD and thermal analysis for framework characterization?
X-ray diffraction and thermal analysis answer different but complementary research questions.
| Technique | Main research question | Information provided |
|---|---|---|
| Powder XRD | What is the crystalline structure of the material? | Phase identification, crystallographic information and structural refinement |
| TGA / STA | How does the material respond to temperature and atmosphere? | Mass changes, thermal behaviour and gas uptake under defined conditions |
| Sample observation STA | Are mass changes accompanied by visible changes? | Correlation of gravimetric data and sample appearance |
Together, these techniques can support a more comprehensive framework characterization workflow.
XRD provides structural information, while STA helps researchers investigate material behaviour under controlled experimental conditions.
For COF research, the suitability of these methods depends on the material’s crystallinity, stability and experimental objectives. The two examples presented here specifically investigate MOFs.
These techniques may also be complemented by other analyses, such as gas sorption measurements for specific surface area and pore-size characterization.
Optimize your MOF and COF characterization workflow with AddSPX
Whether you are developing new porous materials or investigating the performance of existing frameworks, selecting the right analytical techniques is an important step.
At AddSPX, we support researchers with advanced analytical instrumentation, including X-ray diffraction and thermal analysis solutions from Rigaku in Belgium, Luxembourg and The Netherlands.
Would you like to explore which instruments and configurations best suit your research?
Contact our product specialist to discuss your samples, analytical requirements and characterization challenges.
Frequently asked questions about MOF and COF characterization
Can powder XRD determine the crystal structure of a MOF?
Yes. Powder XRD can support crystal structure determination when suitable single crystals are unavailable. Success depends on diffraction data quality, peak overlap and structural complexity. Rigaku’s Cu-MOF application demonstrates ab-initio structure determination using a benchtop MiniFlex 600.
How can CO₂ adsorption in MOFs be measured?
CO₂ adsorption can be investigated through gravimetric measurements under controlled gas atmospheres. For example, thermogravimetric analysis measures mass changes associated with adsorption and desorption. The HKUST-1 study demonstrated an 11.7% mass increase following CO₂ exposure under the specified experimental conditions.
Can XRD and thermal analysis also be used for COFs?
Yes. Powder XRD is relevant for investigating crystalline COF structures and phases. Thermal analysis can help characterize thermal stability and, with appropriate methods and configurations, mass changes associated with gas exposure. Method suitability must be assessed for each material.
Does thermal analysis replace gas sorption measurements?
Not necessarily. Gravimetric thermal analysis can quantify mass changes during gas exposure under defined conditions. Dedicated gas sorption methods may still be required to determine BET surface area, pore-size distributions or adsorption isotherms.